Medtronic 2090 Carelink Programmer, 29901 Encore Programmer, Carelink Network Software 2491 (17 Jan 2019)

Update required to display correct longevity estimate for Micra TPS

Device: Programmer Software Update supporting the Micra Transcatheter Pacing System (TPS):  Medtronic 2090 Carelink Programmer, 29901 Encore Programmer, Carelink Network Application software 2491

ANZCDACC Product Defect Correction Notice 17th Jan 2019

ARTG: 116038 & 213024

TGA Reference: RC-2019-RN-00079-1

Download pdf

Medtronic Adapta DR (Jan 2019)

If error occurs, device will withhold pacing until a VS detected

ANZCDACC Advisory Notice 24th January 2019

 Download as pdf

Device: Medtronic Adapta DR: Model numbers affected

Affected Model numbers ARTG
ADDR01 125076
ADDR03 125077
ADDR06 125078
ADDRL1 125084
ADDRS1 125085
ADVDD01 125080

Not all Adapta devices are affected by this recall. Patients and clinicians may determine if a specific device is affected by looking up the serial number of Medtronic’s Product Performance website.

TGA Reference: RC-2019-RN-00080-1

Boston Scientific VALITUDE™ CRT-P, VISIONIST™ CRT-P, ACCOLADE™ Pacemakers,  PROPONENT™ Pacemakers (24 Dec 2017)

Intermittent over sensing of Minute Ventilation sensor causing periods of pacing inhibition.

ANZCDACC Advisory Notice 24th December 2017

Device: Boston Scientific VALITUDE™ CRT-P Model U128, VISIONIST™ CRT-P Models U225, U226, U228, ACCOLADE™ Pacemakers Models L310, L311, L331,  PROPONENT™ Pacemakers Models L210, L211, L231

TGA Reference:  RC-2017-RN01509-1

Download as pdf

Boston Scientific: VALITUDE™ CRT-P, RESONATE™ CRT-D, MOMENTUM™ CRT-D, AUTOGEN™ CRT-D, VISIONIST™ CRT-P, DYNAGEN™ CRT-D(24Dec2017)

Device: Boston Scientific VALITUDE™ CRT-P Model U128, RESONATE™ CRT-D Model G447, MOMENTUM™ CRT-D Model G124, G125, G126, G128, AUTOGEN™ CRT-D Models G172, G173, G175, G177, G179 VISIONIST™ CRT-P Models U225, U226, U228, DYNAGEN™ CRT-D Models G151, G156, G158

TGA Reference: RC-2017-RN-01510-1

download pdf 